Agilent Technologies 8566B Operating And Programming Manual page 189

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

DSPLY
Display
DSPLY
Item
INTEGER
INTEGER
VARIABLE
IDENTIFIER
The DSPLY command displays the value of a variable anywhere on the spectrum analyzer display.
Field width specifies the number of digits displayed, including sign and decimal point. Places to the right of the
decimal point are limited by decimal places. For example, the number 123.45 has a field of 7, and 2 decimal
places.
Use the DA, PU, PD, and PA commands to position the variable on the screen.
80 Programming
Description/Default
Specifies number of digits
decimal point.
Specifies number of digits to
Alpha character. User-defined
statement.
variable identifier, such as
CF or MA.
Trace element, such as TRA [ 101
F I E L D
DECIMAL
Range Restriction
0 to 9
AA-22 and _
2-12 characters required.