Instrument Preset
The instrument preset command, IP executes the following commands:
CLRW A (Al)
BLANK B (B4)
CR
CA
c s
CT
c v
AMB OFF (Cl)
FA
FB
HD
VAVG OFF (KSH)
DET NRM (KSa)
MKNOISE OFF (KSL)
DET NRM (KSa)
GRAT ON (KSn)
LG
MKTRACK OFF (MT@
MKOFF (Ml)
CONTS (Sl)
THE OFF (TO)
TM FREE (Tl)
TDF P (03)
DA
PD
R3
MKPZ 6dB
MDS W
DISPOSE ONEOS
DISPOSE ONSWP
DISPOSE TRMATH
MKPAUSE OFF
In addition, IP re-assigns user-defined variables to their initial values, specified by the VARDEF command.
Clears and writes trace A.
Blanks trace B .
Couples resolution bandwidth.
Couples input attenuation.
Couples step size.
Couples sweep time.
Couples video bandwidth.
Turns off A-B mode.
Sets start frequency
Sets top frequency.
Hold
Selects dBm amplitude units.
Turns off video averaging.
Selects normal detection mode.
Turns off noise markers.
Selects normal detection mode.
Turns on graticule .
Turns on characters.
Selects lOdB/DIV log scale.
Turns off marker tracking.
Turns off markers.
Selects continuous sweep mode.
Turns off threshold.
Selects free run trigger.
Selects 03 output format.
Selects 3072 as the current address.
Selects normal display size.
Puts pen down at current address.
Allows SRQ 110.
Selects data size of one word, which is two 8-bit bytes.
Erases command list associated with the end of the sweep. (See ONEOS.)
Erases command list associated with the beginning of the sweep. (See ONSWF?!)
Erases command list associated with the end of the sweep. (See TRMATH.)
Turns off marker pause mode.