Agilent Technologies 8566B Operating And Programming Manual page 490

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

ADVANCED DISPLAY PROGRAMMING
The display can now be modified by adding various auxiliary functions to the existing control words.
Brighten the "INPUT" term by adding 128 (brt) to the label address 3089 (1025 + 128 = 1153).
O U T P U T 718;"1153, 7 3 , 7 8 , 80, 85, 8 4 , 1028:"
The label "INPUT" can be made to blink by adding blink on (bk on) and blink off (bk of) words before and after the
"INPUT" label.
OUTPUT 718;"1025,17, 73, 78, 80' 85, 84' 18' 1028:"
Alternately, line 7 could have been replaced with the following lines:
61
OUTPUT 718:"DTB"
OUTPUT 718
1 6 A p p e n d i x
USING "K,B,K,B,K