The KSc command adds trace A to trace B, point by point, and sends the result to trace A. Thus, KSc can restore
the original trace after an A - minus - B function (AMB) is executed.
To successfully add all trace elements, place trace A in VIEW or BLANK display mode before executing KSc.
10
ASSIGN @Sa TO 718
OUTPUT @Sa;"IP,LF;"
20
30
OUTPUT @Sa;"CFlOOMZ;SP2MZ;"
40
OUTPUT @Sa; "A4; "
50
OUTPUT @Sa;"Bl;CF2OOMZ;"
OUTPUT @a; 24; "
60
70
OUTPUT @Sa;"A3;B3;"
80
OUTPUT @Sa; "KSc; "
E N D
90
Line 20:
Presets the instrument.
Line 30:
Sets trace A to 100 MHz center frequency with 2 MHz frequency span.
Line 40:
Blanks trace A.
Selects trace B and sets center frequency to 200 MHz.
Line 50:
Blanks trace B.
Line 60:
Line 70:
Views trace A and trace B.
Combines the amplitude of trace B with trace A and displays this combination as trace A.
Line 80:
The functions of the KSc and APB commands are identical.
The KSc function and the front-panel (,*,, b are identical.
168 Programming
A + B - > A