Marker Noise On
The KSM command displays the RMS noise level at the marker. The RMS value is normalized to a 1 Hz band-
width.
The KSM command averages the amplitude of 32 elements about the location of the marker, in the frequency or
time scale. The average value is converted to the current reference level unit (dBm, dBmV dBuV, or volts).
The noise level function measures accurately to within 10 dB of the analyzer's own internal noise level. The
readout resolution is + - 0.1 dB.
OUTPUT 718;"KSM;"
The functions of the KSM command and the [,,,, IIM*lll keys are identical. See also MKNOISE and KSO.