Agilent Technologies 8566B Operating And Programming Manual page 443

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

T D F
Trace Data Format
The TDF command formats trace information for return to the controller.
OUTPUT 718; "TDF B;"
Specifying M enables the 01 format and returns values in display units, from 0 to 1001.
Specifying P enables the 03 format and returns absolute measurement values, such as dBm or Hz.
Specifying A returns data as an A-block data field. The MDS command determines whether data comprises
one or two 8bit bytes. (See MDS.)
Specifying I returns data as an I-block data field. The MDS command determines whether data comprises
one or two &bit bytes. (See MDS .)
Specifying B enables the 02 or 04 format. The MDS command determines whether data comprises one or
two 8-bit bytes.
See the 01,02,03, and 04 FORMAT commands.