The PA command specifies in display units a vector location on the CRT relative to display reference coordinates
command is in effect. If the pen-up (PU) command is in effect, the vector does not appear of the CRT A sample
program using the PA command is shown below.
ASSIGN @Sa TO718
10
20
OUTPUT @Sa;"IP;A4;KSm;KSo;"
30
OUTPUT @Sa;"DB;PU;"
40
50
80
E N D
Line 20:
Presets the analyzer and clears the display
Line 30:
Specifies the full CRT display size. The pen-up command prevents the initial vector (to point
700,500)
Line 40:
Specifies the starting point of the rectangle to be drawn by the program (coordinates 700,500).
The PD (pen-down) command causes a vector to be drawn on the CRT from the starting point
coordinates to the next set of coordinates (900,500) specified in the program.
Line 50:
Plots the remainder of the rectangle on the CRT The pen-down command remains in effect.
PA
Description/Default
Represents x,y coordinates of vector
endpoint(s), in display units.
from being drawn.
,-Y POSITION7
integer
Range Restriction
Programming 285
PA
Plot Absolute