Agilent Technologies 8566B Operating And Programming Manual page 240

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

The IDSTAT command returns a 1,8, or - 1, representing the completion status of the signal identifier routine, to
the controller.
The 1 means the signal was found, indicating signal identification routine identifies signal and moves signal to
center screen. The pI means signal was not found, analyzer recalled register 7. The - 1 means the signal was
found, indicating signal identification routine found signal but did not move signal to center screen because of a
harmonic lock (see HNLOCK) .
10 OUTPUT 718;"IDSTAT?;"
20 ENTER 718;N
30 PRINT N
4 0
E N D
See SIGID or KSv.
Signal Identifier Status
Programming 131