The BTC command transfers trace B to trace C.
Note that trace C is not a swept, active function. Therefore, transfer trace information to trace C as follows:
Select single sweep mode (S2).
2.
Select desired analyzer settings.
3.
Take one complete sweep (TS) .
4. Transfer data
This procedure ensures that the current settings of the analyzer are reflected in the transferred data.
10
OUTPUT 718;"IP;TS;SNGLS;AB;"
20
30
OUTPUT 718;"BTC;KSj"
31
LOCAL 718
40
E N D
When transferring trace data from one trace to another, only the trace information from 1001 display memory
addresses is transferred out of the total 1024 available display memory addresses. Information in address 1024
and addresses 2026 through 2047 is not transferred. (Addresses 2026 through 2047 are usually used for custom
graphics.)
The functions of the BTC and KS1 commands are identical.
Transfer B to C
P r o g r a m m i n g 4 3