Agilent Technologies 8566B Operating And Programming Manual page 525

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

APPENDIX I
Examples that are least likely to occur are GRAT, PDA4, and PDFA because, in these sets of commands, the
second command will not typically follow the first command. However, all examples have the potential to cause
problems, because they don't follow "B" syntax requirements. The HP 8566B interprets "A" software written like
the above examples as invalid commands. As a result, the commands are not executed and an HP-IB command
error should appear on the analyzer CRT Fortunately, this command error can be used as a method of finding
software errors.
OPERATING DIFFERENCES
Appendix 5 1

Advertisement

Table of Contents
loading

Table of Contents