Agilent Technologies 8566B Operating And Programming Manual page 310

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

Lines 10 to 100:
Sends the content of trace memory to the controller. Refer to the description of the KS123
mnemonic for a complete explanation of these lines.
Lines 120 to 150:
Erases trace A, B, and C memories and blanks the annotation and graticule.
Sends the KS39 command and the display memory address to the analyzer. The USING
Line 170:
part of the OUTPUT statement formats the controller to send the KS as a compact field,
the 39 as a single binary byte, and the PI (display address) as a two byte binary word, the #
sign suppresses the trailing CR/LF so it will not be send as part of the display memory
data.
Line 180:
Sends the display memory data contained in array M$ to the analyzer and terminates the
KS39 command with a 32. The USING part of the OUTPUT statement formats the
controller to end the contents of the array as eight strings and the 32 as a single binary
byte.
Al sets trace A to the clear-write mode. HD clears the active function block of the display,
Line 190:
which contained a display address.
The KS39 command cannot be executed from the front panel.
This is the decimal ASCII equivalent and is transmitted to the analyzer as a single &bit byte.
l
(Continued) KS39