Navair 17-20VQ-05 Technical Manual page 99

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Hewlett–Packard 5306A Multimeter/Counter*
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range
4.1.9
100 V
"
100 V
"
1000 V
TI 10 V Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating–Voltage Measurements
TI Range
4.2.2
10 V
"
10 V
"
10 V
"
100 V
"
100 V
"
1000 V
*The TI calibration is in two parts: First, calibrate the counter function in accordance
with VF–06; then, calibrate the Multimeter function with this procedure, VQ–05.
**The TI may not indicate this value; however, an overrange indication is considered acceptable.
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x .9
9.0000
Calibrator
(V)
10 x .1
10.000
100
100.000
100
100.00
V)
700
300
Calibrator
(V)
400 Hz–1 V
1.0000
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
400 Hz–10 V
10.000
400 Hz–100 V
100.000
400 Hz–100 V
100.00
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(V)
ck ( )
( V)
(V)
ck ( )
ck ( )
ck ( )
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.0003 to
0.0003
0.9994
to
1.0006
1.9991
to
2.0009
2.9988
to
3.0012
3.9985
to
4.0015
4.9982
to
5.0018
5.9979
to
6.0021
6.9976
to
7.0024
7.9973
to
8.0027
8.9970
to
9.0030
8.9970
to
9.0030
(V)
9.994
to
10.006
99.967
to
100.033**
99.60
to
100.40
( V)
600
to
800
200
to
400
(V)
0.9882
to
1.0118
9.9000
to
10.1000**
9.8920
to
10.1080**
9.800
to
10.200
98.450
to
101.550**
98.00
to
102.00
Page 1 of 2
115