Navair 17-20VQ-05 Technical Manual page 112

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8600A Digital Multimeter (Including Options 001 and/or 002)
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
4.1.6
–20 V
4.1.9
2 V
"
200 V
"
1200 V
TI Range Calibrator
4.1.9
200 mV
TI 200 mV Range
4.1.13
10 V LSD
4.1.14
10 V LSD
4.2
Alternating–Voltage Measurements
TI Range
4.2.2
2 V
"
20 V
"
20 V
"
20 V
"
200 V
"
200 V
"
200 V
"
1200 V
"
1200 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
1 V x 1
1.0000
100 V
100.00
100 V
100.0
mV)
1 V x .1
100.00
V)
520
480
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
4 kHz–10 V
10.000
50 kHz–10 V
10.000
50 kHz–10 V
10.00
4 kHz–10 V
10.00
400 Hz–100 V
100.00
400 Hz–100 V
100.0
4 kHz–10 V
10.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(mV)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.002
to
0.002
0.998
to
1.002
1.997
to
2.003
2.997
to
3.003
3.996
to
4.004
4.996
to
5.004
5.996
to
6.004
6.995
to
7.005
7.995
to
8.005
8.994
to
9.006
9.994
to
10.006
0.9994
to
1.0006
99.94
to
100.06
99.8
to
100.2
(mV)
99.91
to
100.09
( V)
510 to 530
470
to
490
(V)
0.9966
to
1.0034
9.966
to
10.034
9.966
to
10.034
9.929
to
10.071
9.87
to
10.13
9.93
to
10.07
99.66
to
100.34
99.2
to
100.8
9.5
to
10.5
Page 1 of 2
129

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents