CALIBRATION CHECKLIST
TEST INST (S)
Cimron DMM 40 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range Calibrator
4.1.9
1 V
"
100 V
"
1000 V
TI 0.1 volt range
4.1.13
10 V LSD
4.1.14
10 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
1 V
"
10 V
"
10 V
"
10 V
"
100 V
"
100 V
"
100 V
"
1000 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
V)
1 V x 1
1.0000
100 V
100.00
100 V
100.0
V)
520
480
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
4 kHz–10 V
10.000
50 kHz–10 V
100.00
400 Hz–100 V
100.00
4 kHz–10 V
10.00
50 kHz–10 V
10.00
400 Hz–100 V
1000.0
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(V)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(V)
–0.001
to
0.001
0.999
to
1.001
1.999
to
2.001
2.999
to
3.001
3.999
to
4.001
4.998
to
5.002
5.998
to
6.002
6.998
to
7.002
7.998
to
8.002
8.998
to
9.002
9.997
to
10.003
(V)
0.9997
to
1.0003
99.97
to
100.03
99.9
to
100.1
( V)
510
to
530
470
to
490
(V)
0.9989
to
1.0011
9.989
to
10.011
9.989
to
10.011
9.905
to
10.095
99.89
to
100.11
9.98
to
10.02
9.86
to
10.14
99.8
to
100.2
Page 1 of 2
67