Navair 17-20VQ-05 Technical Manual page 46

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Systron Donner 9000, 9015, or 9025 Digital Volt/Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
4.3.2
1 M
"
10 M
Affix the Systron Donner 9025 Digital Multimeter with a special calibration label stating:
"This TI has been calibrated up to 50 kHz only."
54
MFG.
NOMINAL
(3)
(k )
range
1.000
range
10.00
range
100.0
(M )
range
1.000
1 M
1.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(k )
0.998
to
1.002
9.98
to
10.02
99.8
to
100.2
(M )
0.998
to
1.002
0.99
to
1.01
Page 2 of 2