NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8010A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
200
4.3.2
2 k
"
20 k
"
200 k
TI Range
4.3.2
2000 M
"
20 M
Affix a special calibration label stating the following:
1. The V ac frequency response has been calibrated only at: 400 Hz,
4 kHz, and 50 kHz.
2. The AC/DC current functions have not been tested.
3. The 200 mV AC range has not been calibrated.
154
MFG.
NOMINAL
(3)
Calibrator
(k )
1 k
1.000
10 k
10.00
100.0 k
100.0
Calibrator
(M )
1.000
1.000
10 M
10.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
(7)
(k )
to .1003
0.997
to 1.003
9.97
to 10.03
99.7
to 100.3
(M )
0.994
to 1.006
9.94
to 10.06
Page 2 of 2