CALIBRATION CHECKLIST
TEST INST (S)
Dana 4700 or 4700A Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range
4.1.9
1 V
"
100 V
"
1000 V
TI 0.1 volt range
4.1.13
10 V LSD
4.1.14
10 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
1 V
"
10 V
"
100 V
"
1000 V
"
10 V
"
10 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x .9
9.000
Calibrator
(V)
10 V x 1
1.0000
100 V
100.00
100 V
100.0
V)
520
480
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
400 Hz–100 V
100.00
400 Hz–100 V
100.0
4 kHz–10 V
10.000
50 kHz–10 V
10.000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(V)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.001
to
0.001
0.999
to
1.001
1.998
to
2.002
2.998
to
3.002
3.998
to
4.002
4.997
to
5.003
5.997
to
6.003
6.997
to
7.003
7.997
to
8.003
8.996
to
9.004
9.996
to
9.004
(V)
0.9996 to
1.0004
99.96
to 100.04
99.9
to 100.1
( V)
510
to 530
470
to 490
(V)
0.9988 to
1.0012
9.998
to
10.012
99.88
to 100.12
99.7
to 100.3
9.988
to
10.012
9.896
to
10.104
Page 1 of 2