Navair 17-20VQ-05 Technical Manual page 47

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Systron Donner 9310, 9320, 9330, 9340, or 9340–5B Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range Calibrator
4.1.9
1000 mV
TI Range
4.1.9
1000 V
TI 100 mV Range
4.1.13
10 V LSD
4.1.14
10 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
1 V
"
10 V
"
100 V
"
1000 V
"
10 V
Range
4.2.2
1000 V
"
10 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
10 V x 1
10.000
mV)
10 V x 1
1000.0
Calibrator
(V)
100 V
100.00
V)
520
480
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
400 Hz–100 V
100.00
400 Hz–100 V
100.0
4 kHz–10 V
10.000
Freq.
(V)
4 kHz
10.0
50 kHz
10.000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(mV)
(V)
( V)
(V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.003 to
0.003
0.997
to
1.003
1.997
to
2.003
2.997
to
3.003
3.997
to
4.003
4.996
to
5.004
5.996
to
6.004
6.996
to
7.004
7.996
to
8.004
8.996
to
9.004
9.996
to
10.004
9.996
to
10.004
(mV)
999.6
to 1000.4
(V)
99.7
to
100.3
( V)
510
to
530
470
to
490
(V)
0.9988 to
1.0012
9.988
to
10.012
99.98
to
100.12
99.7
to
100.3
9.949
to
10.051
(V)
9.8
to
10.2
9.930
to
10.070
Page 1 of 2