Navair 17-20VQ-05 Technical Manual page 17

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Hewlett–Packard 3490A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating:
1. The 0.1 k range has not been calibrated.
2. All other resistance ranges have been calibrated to
3. The V ac function of the TI has been calibrated up to 50 kHz only.
20
MFG.
NOMINAL
(3)
(k )
range
1.00000
range
10.0000
range
100.000
range
1000.00
range
10000.0
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(k )
0.2% of range.
SER. NO.
CALIBRATION TOLERANCES
(7)
(k )
0.99800 to
1.00200
9.9800 to
10.0200
99.800
to
100.200
998.00
to 1002.00
9980.0
to 10020.0
Page 2 of 2