CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8000A/BU Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
4.1.6
–20 V
4.1.9
2 V
"
200 V
"
1200 V
TI Range Calibrator
4.1.9
200 mV
TI 200 mV Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating–Voltage Measurements
TI Range
4.2.2
2 V
"
20 V
"
20 V
"
20 V
"
200 V
"
200 V
"
1200 V
"
1200 V
"
1200 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.00
10 V x .1
1.00
10 V x .2
2.00
10 V x .3
3.00
10 V x .4
4.00
10 V x .5
5.00
10 V x .6
6.00
10 V x .7
7.00
10 V x .8
8.00
10 V x .9
9.00
10 V x 1
10.00
1 V x 1
1.000
100 V
100.0
100 V
100
mV)
1 V x .1
100.0
V)
700
300
Calibrator
(V)
400 Hz–1 V
1.000
400 Hz–10 V
10.00
4 kHz–10 V
10.00
50 kHz–10 V
10.00
4 kHz–10 V
10.0
400 Hz–100 V
100.0
400 Hz–100 V
100
4 kHz–10 V
10
50 kHz–10 V
10
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(mV)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.01
to
0.01
0.99
to
1.01
1.99
to
2.01
2.99
to
3.01
3.99
to
4.01
4.98
to
5.02
5.98
to
6.02
6.98
to
7.02
7.98
to
8.02
8.98
to
9.02
9.98
to
10.02
0.998
to
1.002
99.8
to
100.2
99
to
101
(mV)
99.8
to
100.2
( V)
600
to
800
200
to
400
(V)
0.993
to
1.007
9.93
to
10.07
9.93
to
10.07
9.68
to
10.32
9.7
to
10.3
99.3
to
100.7
97
to
103
8
to
12
8
to
12
Page 1 of 2
149