CALIBRATION CHECKLIST
TEST INST (S)
Hewlett–Packard 2402A with Options 02 and/or 03 Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (03)
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
4.3.2
10 M
Affix a special calibration label stating:
1. The resistance ranges have been calibrated to
2. The 100 mV V dc range has not been calibrated.
3. The V ac function of the TI has been calibrated up to 50 kHz only.
12
MFG.
NOMINAL
(3)
(k )
range
1.00000
range
10.0000
range
100.000
range
1000.00
(M )
range
10.0000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL (6)
(4)
(5)
(k )
(M )
0.2% of range.
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(k )
0.99800 to
1.00200
9.9800
to
10.0200
99.800
to 100.200
998.00
to 1002.00
(M )
9.9800
to
10.0200
Page 2 of 2