Navair 17-20VQ-05 Technical Manual page 54

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8300A Digital Voltmeter with Options 001 and/or 002
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
+10 V
TI Range Calibrator
4.1.9
1000 mV
TI Range
4.1.9
100 V
"
1000 V
TI 100 mV Range
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements (Options 01)
Range
4.2.2
1 V
"
10 V
"
10 V
"
10 V
"
100 V
"
100 V
"
100 V
"
1000 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x 1
10.0000
10 V x 1
–10.0000
mV)
1 x 1
1000.00
Calibrator
(V)
100 V
100.000
100 V
100.00
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.00000
400 Hz–10 V
10.0000
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
50 kHz–10 V
10.000
4 kHz–10 V
10.000
400 Hz–100 V
100.000
400 Hz–100 V
100.000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(mV)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.0001 to
0.0001
0.9993
to
1.0007
1.9992
to
2.0008
2.9990
to
3.0010
3.9989
to
4.0011
4.9987
to
5.0013
5.9986
to
6.0014
6.9984
to
7.0016
7.9983
to
8.0017
8.9981
to
9.0019
9.9980
to
10.0020
–0.9980 to
–10.0020
(mV)
999.83
to
1000.17
(V)
99.865
to
100.135
99.93
to
100.07
( V)
501
to
503
497
to
499
(V)
0.99860 to
1.00140
9.9860
to
10.0140
9.9860
to
10.0140
9.9760
to
10.0240
9.940
to
10.060
9.950
to
10.050
99.860
to
100.140
99.50
to
100.50
Page 1 of 2
63