NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Dana 4700 or 4700A Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI & Calibrator
4.3
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating:
1. The 0.1 V dc and the 0.1 k range have not been calibrated.
2. All other resistance ranges have been calibrated to
3. The V ac function of the TI has been calibrated up to 50 kHz only.
26
MFG.
NOMINAL
(3)
(k )
range
1.0000
range
10.000
range
100.00
range
1000.0
range
10000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(k )
0.2% of range.
SER. NO.
CALIBRATION TOLERANCES
(7)
(k )
9.9980 to
1.0020
9.980
to
10.020
99.80
to 100.20
998.0
to 1002.0
9980
to10020
Page 2 of 2