Equipment Requirements - Navair 17-20VQ-05 Technical Manual

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
TI
Characteristics
Alternating voltage
measurement
Resistance
Minimum use specifications are the principal parameters required for performance of the calibration, and are
included to assist in the selection of alternate equipment, which may be used at the discretion of the using
laboratory. Satisfactory performance of alternate items shall be verified prior to use. All applicable
equipment must bear evidence of current calibration.
The instruments utilized in this procedure were selected from those known to be available at Navy
calibration facilities, and the listing by make or model number carries no implication of preference,
recommendation, or approval for use by other agencies. It is recognized that equivalent equipment produced
by other manufacturers may be capable of equally satisfactory performance in this procedure.
If the TI has an electronic counter function, calibration is in two parts: First, calibrate the TI electronic
counter function in accordance with NAVAIR 17–20VF–06; then calibrate the multimeter function with this
procedure.
Item
2.1 Portable meter
calibrator
(calibrator)
2
Performance
Specifications
Range: 1, 10, 100, and 1000 V
Tolerance:
1 V at 400 Hz;
0.15% iv;
10 V at 400 Hz or 4 kHz,
0.12% iv;
10 V at 50 kHz;
0.3% iv;
100 V or 1000 V at 400 Hz,
0.18% iv
Range: 1 k through 10 M
Tolerance: Assigned
0.2%.
10
and 100
ranges are not
tested.
SECTION 2

EQUIPMENT REQUIREMENTS

Table 2. Equipment Requirements
Minimum Use
Specifications
DCV range: 0 to 999 V, 1 V, 10 V,
and 100 V
Uncertainty: 0 to 999 V:
1 V, 10 V, and 100 V:
0.003% iv
or 30 V, whichever is greater
ACV range: 1 V, 10 V, and 100 V
Uncertainty: 1 V at 400 Hz:
10 V at 400 Hz and 4 kHz:
10 V at 50 kHz:
0.1%
100 V at 400 Hz:
0.06%
Resistance range: 1k , 1 M , and 10 M
Uncertainty: 1 k and 1 M :
10 M :
0.075%
The TI indication is
compared to the calibrator
V ac output.
The TI indication is
compared to the calibrator
resistance output
NOTE
Fluke 515A (no substitute)
2.0 V
0.05%
0.04%
0.015%
Test
Method
Calibration
Equipment