Navair 17-20VQ-05 Technical Manual page 91

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Vidar 500 Digital Voltmeter Counter*
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
4.1.6
–10 V
4.1.9
1 V
"
100 V
"
1000 V
TI Range Calibrator
"
100 mV 1 V x .1
TI 100 mV Range
4.1.13
100 V LSD
4.1.14
100 V LSD
*Calibration is in two parts: First, calibrate the TI electronic counter function in accordance
with VF–06; then, calibrate the voltmeter function with this procedure, VQ–05.
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.00
10 V x .1
1.00
10 V x .2
2.00
10 V x .3
3.00
10 V x .4
4.00
10 V x .5
5.00
10 V x .6
6.00
10 V x .7
7.00
10 V x .8
8.00
10 V x .9
9.00
10 V x 1
10.00
1 V x 1
1.000
100 V
100.0
100 V
100
mV)
100.0
V)
700
300
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(V)
(mV)
( V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(V)
–0.01
to
0.01
0.99
to
1.01
1.99
to
2.01
2.99
to
3.01
3.99
to
4.01
4.98
to
5.02
5.98
to
6.02
6.98
to
7.02
7.98
to
8.02
8.98
to
9.02
9.98
to
10.02
0.998
to
1.002
99.8
to
100.2
99
to
100
(mV)
99.8
to
100.2
( V)
600
to
800
200
to
400
Page 1 of 1
(106 blank) 105