Navair 17-20VQ-05 Technical Manual page 93

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Data Precision 245 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range*
4.1.3
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
4.1.6
–10 V
4.1.9
1 V
"
100 V
"
1000 V
TI 1 V Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating–Voltage Measurements
TI Range
4.2.2
1 V
"
10 V
"
10 V
"
10 V
"
100 V
"
100 V
"
1000 V
"
1000 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.0000
10 V x .2
2.00
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
1 V x 1
1.00000
100 V
100.00
100 V
100.0
V)
700
300
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
4 kHz–10 V
10.000
50 kHz–10 V
10.000
50 kHz–10 V
10.00
400 Hz–100 V
100.00
400 Hz–100 V
100.0
50 kHz–10 V
10.0
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(V)
–0.001
to
0.001
0.998
to
1.002
1.998
to
2.002
2.997
to
3.003
3.997
to
4.003
4.996
to
5.004
5.996
to
6.004
6.995
to
7.005
7.995
to
8.005
8.994
to
9.006
9.994
to
10.006
0.9994
to
1.0006
99.94
to
100.06
99.8
to
100.2
( V)
600
to
800
200
to
400
(V)
0.9990 to
1.0010
9.990
to
10.010
9.938
to
10.062
9.895
to
10.105
9.85
to
10.15
99.90
to
100.10
99.7
to
100.3
9.4
to
10.6
Page 1 of 2
109

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents