Navair 17-20VQ-05 Technical Manual page 62

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
ME–346/ASM–308, Cimron 9200B, 9300B, 9400B, or 9500B Digital Voltmeter with
the 6980B AC Converter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range Calibrator
4.1.9
1000 mV
TI Range
4.1.9
100 V
"
100 V
"
1000 V
TI 100 mV Range
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
10 V
"
10 V
"
10 V
"
100 V
"
1000 V
*The TI may not indicate this value; however, overrange indication is considered acceptable.
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x .9
9.0000
mV)
1 V x .9
900.00
Calibrator
(V)
10 V x 1
10.000
100 V
100.000
100 V
100.00
V)
502
498
Calibrator
(V)
4 kHz–10 V
10.0000
400 Hz–10 V
10.0000
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
400 Hz–100 V
100.00
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(mV)
(V)
ck ( )
ck ( )
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.0001 to
0.0001
0.9998
to
1.0002
1.9997
to
2.0003
2.9996
to
3.0004
3.9995
to
4.0005
4.9994
to
5.0006
5.9993
to
6.0007
6.9992
to
7.0008
7.9991
to
8.0009
8.9990
to
9.0010
8.9990
to
9.0010
(mV)
899.82
to
900.18
(V)
9.998
to
10.002
99.989
to
100.011*
99.98
to
100.02
( V)
501
to
503
497
to
499
(V)
9.9950
to
10.0050*
9.9950
to
10.0050*
0.9980
to
1.0020
9.980
to
10.020
99.80
to
100.20
Page 1 of 2
71

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05