NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8400A with Options 001, 002, 005, and/or 007, 8425A, 8425A/AF, or 8425B Digital Voltmeter
with Options 001 and/or 002
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (02)
Range & Calibrator
4.3.2
1000
"
10000
Range & Calibrator
4.3.2
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating:
1. The 0.1 V dc, 10 , and 100 ranges have not been calibrated.
2. All other resistance ranges have been calibrated to +0.2% of range.
3. The V ac function has been calibrated up to 50 kHz only.
NOTE: The Fluke 8425 A/AF does not have the 10 & 100 ranges.
50
MFG.
MODEL
MEASURED VALUES
NOMINAL
FIRST RUN
(3)
(4)
)
( )
1000.00
10000.0
k )
(k )
100.000
1000.00
10000.0
SER. NO.
OUT
SECOND RUN
CALIBRATION TOLERANCES
OF
TOL
(5)
(6)
998.00
9980.0
99.800
998.00
9980.0
Page 2 of 2
(7)
( )
to 1002.00
to 10020.0
(k )
to 100.200
to 1002.00
to 10020.0