Navair 17-20VQ-05 Technical Manual page 13

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Cimron Digital Multimeter, Sanders P/N 3005896G1 as used with 0Q–10/ALM–64*
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
–10 V
"
TI Range
4.1.9
100 V
"
1000 V
TI 10–volt range
4.1.13
100 V LSD
4.1.14
100 V LSD
*Affix a special calibration label to the TI stating:
"This TI is calibrated in accordance with the usage requirement of 0Q–10/ALM–64".
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x 1
10.0000
10 V x 1
10.0000
Calibrator
(V)
100 V
100.000
100 V
100.00
V)
700
300
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(V)
(V)
( V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.0002
to
0.0002
0.9997
to
1.0003
1.9995
to
2.0005
2.9992
to
3.0008
3.9990
to
4.0010
4.9987
to
5.0013
5.9985
to
6.0015
6.9982
to
7.0018
7.9980
to
8.0020
8.9977
to
9.0023
9.9975
to
10.0025
9.9975
to
10.0025
(V)
99.975
to 100.025
99.97
to 100.03
( V)
600
to
800
200
to
400
Page 1 of 1
(14 blank) 13