Navair 17-20VQ-05 Technical Manual page 104

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Dana 5000 Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
4.1.9
100 mV
"
1 V
"
100 V
"
1000 V
TI 100 mV Range
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements (Model 34 AC Converter)
TI Range
4.2.2
1 V
"
10 V
"
10 V
"
10 V
"
100 V
"
100 V
"
100 V
"
1000 V
"
1000 V
"
1000 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x 1
10.0000
1 V x .1
10.0000
1 V x 1
1.00000
100 V
100.000
100 V
100.00
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.00000
400 Hz–10 V
10.0000
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
50 kHz–10 V
10.000
4 kHz–10 V
10.000
400 Hz–100 V
100.000
400 Hz–100 V
100.00
4 kHz–10 V
10.000
50 kHz–10 V
10.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(V)
–0.0002 to
0.0002
0.9997
to
1.0003
1.9996
to
2.0004
2.9995
to
3.0005
3.9994
to
4.0006
4.9993
to
5.0007
5.9992
to
6.0008
6.9991
to
7.0009
7.9990
to
8.0010
8.9989
to
9.0011
9.9988
to
10.0012
0.099910 to
0.100090
0.99988 to
1.00012
99.988
to
100.012
99.97
to
100.03
( V)
501
to
503
497
to
499
(V)
0.99880 to
1.00120
9.9880
to
10.0120
9.9880
to
10.0120
9.9830
to
10.0170
9.965
to
10.035
9.970
to
10.030
99.880
to
100.120
99.70
to
100.30
9.79
to
10.21
9.78
to
10.22
Page 1 of 2
121

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents