CALIBRATION CHECKLIST
TEST INST (S)
Systron Donner 7000 or 7005A Digital Volt/Multimeter with Options 02, 03, and/or 07
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range Calibrator
4.1.9
100 V
"
1000 V
TI 100 mV Range
4.1.13
1 V LSD
4.1.14
1 V LSD
TI 1 V Range (TI without Option 07)
4.1.13
10 V LSD
4.1.14
10 V LSD
4.2
Alternating–Voltage Measurements (02)
Range
4.2.2
1 V
"
10 V
"
100 V
"
1000 V
"
10 V
"
10 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x 1
10.0000
10 V x 1
10.0000
V)
100 V
100.000
100 V
100.00
V)
502
498
V)
520
480
Calibrator
(V)
400 Hz–1 V
1.00000
400 Hz–10 V
10.0000
400 Hz–100 V
100.000
400 Hz–100 V
100.00
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(V)
(V)
( V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.0004
to
0.0004
0.9995
to
1.0005
1.9994
to
2.0006
2.9993
to
3.0007
3.9992
to
4.0008
4.9991
to
5.0009
5.9990
to
6.0010
6.9989
to
7.0011
7.9988
to
8.0012
8.9987
to
9.0013
9.9986
to 10.0014
9.9986
to 10.0014
(V)
99.986
to 100.014
99.95
to 100.05
( V)
501
to
503
497
to
499
( V)
510
to
530
470
to
490
(V)
0.99700 to
1.00300
9.9700 to
10.0300
99.700
to 100.300
97.90
to 102.10
9.9700 to
10.0300
9.9000 to
10.1000
Page 1 of 2