Navair 17-20VQ-05 Technical Manual page 128

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8000A/BU Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
200
TI Range
4.3.2
2 k
"
20 k
"
200 k
"
2000 k
TI Range
"
20 M
Affix a special calibration label stating:
"The AC/DC current functions have not been calibrated: also, the 200
range has been calibrated to +0.05% of indication + 1 digit."
150
MFG.
NOMINAL
(3)
Calibrator
( )
Calibrator
(k
1 k
1.000
10 k
10.00
100 k
100.0
1 M
1000
Calibrator
(M )
10 M
10.00
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
( )
(k )
(M )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
( )
to
100.6
(k )
0.997
to
1.003
9.97
to
10.03
99.7
to
100.3
997
to
1003
(M )
9.94
to
10.06
Page 2 of 2

Advertisement

Table of Contents
loading