Navair 17-20VQ-05 Technical Manual page 106

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Cimron 6753 or 6753–1424 Digital Multimeter with the 5775 or 5789 Plug in Unit
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
4.1.9
1 V
4.1.9
100 V
"
1000 V
TI 10 mV Range (5789)
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
1 V
"
10 V
"
10 V
"
10 V
"
100 V
"
100 V
"
100 V
"
1000 V
"
1000 V
"
1000 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x 1
10.0000
1 V x 1
1.00000
100 V
100.000
100 V
100.000
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.00000
400 Hz–10 V
10.0000
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
50 kHz–10 V
10.000
4 kHz–10 V
10.000
400 Hz–100 V
100.000
400 Hz–100 V
100.00
4 kHz–10 V
10.00
50 kHz–10 V
10.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(V)
–0.0002 to
0.0002
0.9997
to
1.0003
1.9996
to
2.0004
2.9995
to
3.0005
3.9994
to
4.0006
4.9993
to
5.0007
5.9992
to
6.0008
6.9991
to
7.0009
7.9990
to
8.0010
8.9989
to
9.0011
9.9988
to
10.0012
0.99985 to
1.00015
99.988
to
100.012
99.97
to
100.03
( V)
501
to
503
497
to
499
(V)
0.99850 to
1.00150
9.9850
to
10.0150
9.9850
to
10.0150
9.9500
to
10.0500
9.930
to
10.070
9.972
to
10.028
99.850
to
100.150
99.72
to
100.28
9.79
to
10.21
9.48
to
10.52
Page 1 of 2
123

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents