Navair 17-20VQ-05 Technical Manual page 115

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8800A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
200
TI Range
4.3.2
2 k
"
20 k
"
200 k
"
2000 k
TI Range
4.3.2
20 M
Affix a special calibration label stating the following:
1. The V ac frequency response has been calibrated only at: 400 Hz, 4 kHz, & 50 kHz.
2. The 200
3. The 2, 20 and 200 k resistance ranges have been calibrated to +(0.196% iv + 0.002% of range).
4. The 200 mV range of the direct voltage function has been calibrated to
+(0.087 + 0.0015% of range).
132
MFG.
NOMINAL
(3)
Calibrator
( )
100
100.00
Calibrator
(k )
1 k
1.00000
10 k
10.0000
100 k
100.000
1 M
1000.00
Calibrator
(M )
10 M
10.0000
resistance range has been calibrated to +(0.62% iv + 0.005% of range).
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
( )
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
( )
99.370
to
100.630
(k )
0.99800 to
1.00200
9.9800
to
10.0200
99.800
to
100.200
998.96
to
1001.04
(M )
9.9594
to
10.0406
Page 2 of 2

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05