NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8110A, 8110AOPT001, 8125A, or 8125AOPT001 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
4.3.2
10 M
Affix a special calibration label stating:
"All resistance ranges have been calibrated to +0.2% of range.
No measurement above 4 kHz has been calibrated."
44
MFG.
NOMINAL
(3)
(k )
range
1.0000
range
10.000
range
100.00
range
1000.0
(M )
10.000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
(M )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.9980
to 1.0020
9.980
to 10.020
99.80
to 100.20
998.0
to 1002.0
(M )
9.980
to 10.020
Page 2 of 2