Navair 17-20VQ-05 Technical Manual page 134

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Beckman 310 or TECH 310 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
4.3.2
TI Range
"
2 k
"
20 k
"
200 k
TI Range
4.3.2
2 k
"
20 M
Affix a special calibration label stating the following:
1. The V ac frequency response has been calibrated only at: 400 Hz and 4 kHz.
2. The AC/DC current functions have not been tested.
3. The 200 mV AC range has not been calibrated.
156
MFG.
NOMINAL
(3)
Calibrator
(k
1 k
1.000
10 k
10.00
100 k
100.0
Calibrator
(M )
1 M
1.000
10 M
10.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
(7)
(k )
0.940 to 1.006
9.94
to 10.06
99.4
to 100.6
(M )
0.994 to 1.006
9.94
to 10.06
Page 2 of 2