NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8120A Digital Voltmeter including Options 001 and/or 002
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
k Ranges:
4.3.2
1 k
"
10 k
"
100 k
4.3.2
10 M
Range
Affix a special calibration label to the following:
1. The TI V ac frequency response has not been calibrated.
2. The 100 mVac range has not been calibrated.
3. The 1 k and 10 M ranges have been calibrated to +0.2% or range.
4. The DC and AC current functions have not been calibrated.
5. The 100 mV V dc range has not been calibrated.
94
MFG.
MODEL
MEASURED VALUES
NOMINAL
FIRST RUN
(3)
(4)
(k )
(k )
1.0000
10.000
100.00
(M )
(M )
10.000
SER. NO.
OUT
SECOND RUN
CALIBRATION TOLERANCES
OF
TOL
(5)
(6)
0.9980 to
9.994
99.94
9.980
Page 2 of 2
(7)
(k )
1.0020
to
10.006
to
100.06
(M )
to
10.020