CALIBRATION CHECKLIST
TEST INST (S)
Cimron 9500B–355 Digital Multimeter with the 6980B AC Converter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range Calibrator
4.1.9
1000 mV
TI Range
4.1.9
100 V
"
100 V
"
1000 V
TI 10 V Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
10 V
"
10 V
"
100 V
"
100 V
"
1000 V
Affix as special calibration label stating:
"The direct–voltage ratio measurement function and the 100 mV dc range have not been calibrated."
*The TI may not indicate this value; however, an overrange indication is considered acceptable.
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x .9
9.0000
mV)
1 x 0.9
900.00
Calibrator
(V)
10 V
10.000
100 V
100.000
100 V
100.00
V)
700
300
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.0000
400 Hz–10 V
10.000
400 Hz–100 V
100.000
400 Hz–100 V
100.00
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(mV)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.0001 to
0.0001
0.9998
to
1.0002
1.9997
to
2.0003
2.9996
to
3.0004
3.9995
to
4.0005
4.9994
to
5.0006
5.9993
to
6.0007
6.9992
to
7.0008
7.9991
to
8.0009
8.9990
to
9.0010
8.9990
to
9.0010
(mV)
899.82
to
900.18
(V)
99.998
to
10.002
99.989
to
100.011*
99.98
to
100.02
V)
600
to
800
200
to
400
(V)
0.9960
to
10.0040
9.9870
to
10.0130*
9.960
to
10.040*
99.870
to
100.130*
99.60
to
100.40
Page 1 of 1