CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8800A/AA Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
20 V
"
20 V
"
20 V
"
20 V
"
20 V
"
20 V
"
20 V
"
20 V
"
20 V
"
20 V
4.1.6
–20 V
4.1.9
2 V
"
200 V
"
1200 V
TI 200 mV Range
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
2 V
"
20 V
"
20 V
"
20 V
"
200 V
"
200 V
"
1200 V
"
1200 V
"
1200 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x 1
10.0000
1 V x 1
1.0000
100 V
100.000
100 V
100.00
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.00000
400 Hz–10 V
10.0000
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
50 kHz–10 V
10.000
400 Hz–100 V
100.000
400 Hz–100 V
100.000
4 kHz–10 V
10.00
50 kHz–10 V
10.00
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.0010 to
0.0010
0.9988
to
1.0012
1.9987
to
2.0013
2.9985
to
3.0015
3.9984
to
4.0016
4.9982
to
5.0018
5.9981
to
6.0019
6.9979
to
7.0021
7.9978
to
8.0022
8.9976
to
9.0024
9.9975
to
10.0025
0.99975 to
1.00025
99.975
to
100.025
99.920
to
100.080
( V)
501
to
503
497
to
499
(V)
0.99820 to
1.00180
9.9820
to
10.0180
9.9820
to
10.0180
9.9420
to
10.0580
9.870
to
10.130
99.820
to
100.180
99.42
to
100.58
9.51
to
10.49
9.47
to
10.53