NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Cimron 76, 7630, 7650, or 7650–969 Digital Multimeter with the 6770 or 6771 AC Converter,
or the 6990 Ohm Converter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (6990 Ohm Converter)
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
4.3.2
10 M
Affix special calibration labels to the TI as follows:
1. If the TI has a ratio measurement function, affix a special calibration
label stating, "The ratio function has not been calibrated."
2. If the TI has a 20 kHz bandwidth, affix a special calibration label
stating, "The frequency response has not been calibrated."
3. If the TI has a resistance – measurement function, affix a special
calibration label stating, "All resistance ranges have been calibrated to
+0.2% of range."
4. If the AC Converter is a model 6771, affix a special calibration label
stating: "The V ac function has been calibrated up to 50 kHz only."
*The TI may not indicate this value; however, an overrange indication is considered acceptable.
92
MFG.
NOMINAL
(3)
(k )
1.0000
10.000
100.00
1000.0
(M )
range
10.000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
ck ( )
ck ( )
ck ( )
ck ( )
(M )
ck ( )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.9980
to
1.0020*
9.9800
to
10.020*
99.800
to
100.20*
998.00
to
1002.0*
(M )
9.9800
to
10.020*
Page 2 of 2