CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8400A with Options 001, 002, 005, and/or 007, 8425A, 8425A/AF, or 8425B Digital Voltmeter
with Options 001 and/or 002
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range Calibrator
4.1.9
1 V
"
100 V
"
1000 V
TI 0.1 Volt Range
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements (01)
Range
4.2.2
1 V
"
10 V
"
100 V
"
1000 V
"
10 V
"
10 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.000
10 V x .3
3.00000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x 1
10.0000
10 V x 1
10.0000
V)
1 V x 1
1.00000
100 V
100.000
100 V
100.00
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.00000
400 Hz–10 V
10.0000
400 Hz–100 V
100.000
400 Hz–100 V
100.00
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.0001
to
0.0001
0.9998
to
1.0002
1.9996
to
2.0004
2.9996
to
3.0004
3.9995
to
4.0005
4.9994
to
5.0006
5.9993
to
6.0007
6.9992
to
7.0008
7.9991
to
8.0009
8.9990
to
9.0010
9.9989
to
10.0011
9.9989
to
10.0011
(V)
0.99978
to 1.00022
99.989
to 100.011
99.98
to 100.02
( V)
501
to 503
497
to 499
(V)
0.99945
to 1.00055
9.9945
to 10.0055
99.945
to 100.055
99.90
to 100.10
9.9945
to 10.0055
9.9495
to
10.0505
Page 1 of 2