CALIBRATION CHECKLIST
TEST INST (S)
Non–Linear System 5005 or 5010 Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range
4.1.9
1 V (5010)
"
100 V
"
1000 V
(5010) 1 volt range
4.1.13
100 V LSD
4.1.14
100 V LSD
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
10 V x 1
10.000
Calibrator
(V)
1 V x 1
1.0000
100 V
100.00
100 V
100.0
V)
700
300
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(V)
(V)
( V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.001
to
0.001
0.999
to
1.001
1.999
to
2.001
2.999
to
3.001
3.999
to
4.001
4.998
to
5.002
5.998
to
6.002
6.998
to
7.002
7.998
to
8.002
8.998
to
9.002
9.998
to
10.002
9.998
to
10.002
(V)
0.9998 to
1.0002
99.98
to 100.02
99.9
to 100.1
( V)
600
to
800
200
to
400
Page 1 of 1