CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8110A, 8110AOPT001, 8125A, or 8125AOPT001 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.3
+10 V
4.1.6
–10 V
TI Range
4.1.9
1 V
"
100 V
"
1000 V
TI Volt Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
1 V
"
10 V
"
100 V
"
1000 V
"
10 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
10 V x 1
10.000
Calibrator
V)
10 V x 1
1.0000
100 V
100.00
100 V
100.0
V)
( V)
700
300
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
400 Hz–100 V
100.00
400 Hz–100 V
100.0
4 kHz–10 V
10.000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.002
to
0.002
0.998
to
1.002
1.998
to
2.002
2.997
to
3.003
3.997
to
4.003
4.997
to
5.003
5.997
to
6.003
6.997
to
7.003
7.996
to
8.004
8.996
to
9.004
9.996
to
10.004
9.996
to
10.004
(V)
0.9996 to
1.0004
99.96
to 100.04
99.8
to 100.2
( V)
600
to 800
200
to 400
(V)
0.9970 to
1.0030
9.970
to 10.030
99.70
to 100.30
99.8
to 100.2
9.970
to 10.030
Page 1 of 2