NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Dana 5500 as used with Mobil Automatic Test Set LN2C, 5600, 5600OPT20, or 5600–S638
Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
k Ranges:
4.3.2
1
"
10
"
100
"
1000
"
10,000
Affix a special calibration label stating:
1. The direct–voltage ratio function and the 100 mV V dc range have
not been calibrated.
2. All resistance ranges have been calibrated to +0.2% of range.
3. The V ac function has been calibrated up to 50 kHz only.
88
MFG.
MODEL
MEASURED VALUES
NOMINAL
FIRST RUN
(3)
(4)
(k )
(k )
1.00000
10.0000
100.000
1000.00
10000.0
SER. NO.
OUT
SECOND RUN
CALIBRATION TOLERANCES
OF
TOL
(5)
(6)
0.99800 to
9.9800
99.800
998.00
998.00
Page 2 of 2
(7)
(k )
1.00200
to
10.0200
to
100.200
to
1002.00
to
1002.00