NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8350A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
4.3.2
10 M
Affix a special calibration label stating:
1. The 100 mV V dc range has not been calibrated.
2. All resistance ranges have been calibrated to +0.2% of range.
3. The V ac function has been calibrated up to 50 kHz only.
58
MFG.
MODEL
MEASURED VALUES
NOMINAL
FIRST RUN
(3)
(4)
(k )
(k )
1.00000
10.0000
100.000
1000.00
(M )
(M )
10.0000
SER. NO.
OUT
SECOND RUN
CALIBRATION TOLERANCES
OF
TOL
(5)
(6)
Assigned: +0.2%
0.99800 to
9.9800
99.800
998.00
9.9800
Page 2 of 2
(7)
(k )
1.00200
to
10.0200
to
100.200
to
1002.00
(m )
to
10.0200