NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
ME–346/ASM–308, Cimron 9200B, 9300B, 9400B, or 9500B Digital Voltmeter with
the 6980B AC Converter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
1 k
"
10 k
"
10 k
"
100 k
"
100 k
TI Range
4.3.2
1 M
"
1 M
"
10 M
"
10 M
Affix a special calibration label stating:
"All resistance ranges have been calibrated to +0.2% of range. The
TI V ac frequency response and the 100 mV V dc range have not
been calibrated."
*The TI may not indicate this value; however, an overrange indication
is considered acceptable.
72
MFG.
NOMINAL
(3)
Calibrator
(k )
1 k
1.00000
1 k
1.0000
10 k
10.0000
10 k
10.000
100 k
100.000
Calibrator
(M )
100 k
0.100000
1 M
1.00000
1 M
1.0000
10 M
10.0000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(k )
ck ( )
ck ( )
ck ( )
(M )
ck ( )
ck ( )
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(k )
0.99800 to
1.00200*
0.9980
to
1.0020
9.9800
to
10.0200*
9.980
to
10.020
99.800
to
100.200*
(M )
0.09980 to
0.10020
0.99800 to
1.00200*
0.99800 to
1.0020
0.9800
to
10.0200*
Page 2 of 2