Navair 17-20VQ-05 Technical Manual page 61

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Cimron DMM 50 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating:
1. The 100
have not been calibrated.
2. The V ac frequency response, and measurements above 50 kHz
have not been calibrated.
3. The 10000 k range has been calibrated up to +0.2% of range.
70
MFG.
NOMINAL
(3)
(k )
1 k
1.00000
10 k
10.0000
100 k
100.000
1000 k
1000.00
10000 k
10000.0
range, 0.1 V dc range, and the V dc ratio function
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(k )
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(k )
0.99847 to
1.00153
9.9847
to
10.0153
99.847
to
100.153
998.47
to
1001.53
9980.0
to
10020.0
Page 2 of 2