CALIBRATION CHECKLIST
TEST INST (S)
Beckman 310 or TECH 310 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
4.1.6
–20 V
TI Range
4.1.9
200 mV
"
2 V
"
200 V
"
1500 V
TI 200 mV Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating–Voltage Measurements
TI Range
4.2.2
2 V
"
20 V
"
20 V
"
200 V
"
200 V
"
1000 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 x 0
0.00
10 x .1
1.00
10 x .3
3.00
10 x .5
5.00
10 x .7
7.00
10 x .9
9.00
10 x .9
–9.00
Calibrator
V dc)
1 V x .1
100.0 mV
1 V x 1
1.000
100 V
100.0
100 V
100.0
V dc)
700
300
Calibrator
(V ac)
400 Hz–1 V
1.000
400 Hz–10 V
10.00
4 kHz–10 V
10.00
4 kHz–10 V
10.00
400 Hz–100 V
100.0
400 Hz–100 V
100.0
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(V)
(V dc)
( V dc)
(V ac)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.01 to
0.01
0.98 to
1.02
2.98 to
3.02
4.97 to
5.03
6.97 to
7.03
8.96 to
9.04
–8.96 to –9.04
(V dc)
99.6to 100.4 mV
0.996to 1.004
99.6to 100.4
99.6to 100.4
( V dc)
600 to 800
200 to 400
(V ac)
0.989to 1.011
9.89to 10.11
9.80to 10.20
9.80to 10.20
98.9to 101.1
98.9to 101.1
Page 1 of 2
155