NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Dana 5000 Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (model 54 ohms Converter)
TI Range
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating the following as applicable:
1. The 100 mV of the direct voltage range has been calibrated to +0.09% of range.
2. If the TI consists of a model 54 ohms Converter, the 0.1 k range has not been
calibrated and all other resistance ranges have been calibrated to +0.2% of range.
3. If the TI consists of an option 64, the AC/DC ratio measurement function has not
been calibrated.
4. If the TI consist of a model 34 AC Converter, the V ac function has been
calibrated up to 50 kHz only.
122
MFG.
NOMINAL
(3)
Calibrator
(k )
1k
1.00000
10k
10.0000
100 k
100.000
1 M
1000.00
10 M
10000.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.99800 to
1.00200
9.9800
to
10.0200
99.800
to
100.200
998.00
to
1002.00
9980.0
to
10020.0
Page 2 of 2