Navair 17-20VQ-05 Technical Manual page 109

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8000A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
200
TI Range
4.3.2
2 k
"
20 k
"
200 k
"
2000 k
TI Range
4.3.2
20 M
Affix a special calibration label stating:
"The AC/DC current functions have not been calibrated; also, the
200
range has been calibrated to +0.5% of indication +1 digit."
126
MFG.
NOMINAL
(3)
Calibrator
( )
100
Calibrator
(k )
1 k
1.000
10 k
10.00
100 k
100.0
1 M
1000
Calibrator
(M )
10 M
10.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
( )
(k )
"
"
(M )
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
( )
to
(k )
0.997
to
1.003
9.97
to
10.03
99.7
to
100.3
997
to
1003
(M )
9.94
to
10.06
Page 2 of 2